m/z
Mass-to-charge ratio used to label ion signals.
Quick reference
Look up a term or check a concise planning note. For a complete lesson, use the subject index.
Glossary
Useful for beginners, but written so experienced users do not have to tolerate oversimplified explanations. Terms are intentionally connected to how they are used in troubleshooting and review.
Mass-to-charge ratio used to label ion signals.
Formation of ions from atoms or molecules. The resulting ion may have a different composition or charge state from the original analyte.
Subsystem that separates ions by their motion behavior linked to m/z.
Hardware that converts ion behavior into electrical signal, through ion impact or induced image current depending on the instrument.
An ion species formed by association (for example protonated, sodiated, or ammoniated molecules) that shifts observed m/z and may alter fragmentation behavior.
Parts-per-million deviation between observed and expected m/z. Useful for mass accuracy monitoring, but not sufficient by itself to confirm identity.
Difference between measured and expected mass value.
Remaining fit error after calibration model correction.
Limits of detection and quantitation describing the lowest detectable or reliably quantifiable signal under defined conditions and criteria.
A measure of separating nearby peaks. Report the criterion and m/z; the FWHM convention uses R = (m/z) / peak width. High resolving power alone does not guarantee low mass error.
Signal range where response remains useful and interpretable.
Background signal level under peaks.
Accuracy concerns agreement with a reference; precision concerns agreement among repeated measurements under stated conditions. Repeatability is one set of conditions. A tightly grouped set can still be systematically offset.
Mass-axis calibration aligns measured m/z positions with reference values. Quantitative calibration separately relates response to concentration.
Adjusting operating parameters for stable and useful signal behavior.
Effects of sample components on analyte response, which may involve sample introduction, ionization, transmission or other processes. Ion suppression in ESI is one example, not the entire definition.
Signal from previous samples contaminating later runs.
A selected substance or element added at a defined amount to help track response variation. Correction is limited to processes after its addition and to effects it follows sufficiently like the analyte; it cannot repair every loss or interference.
A known ion used to monitor or correct mass accuracy during acquisition. It improves stability tracking but does not fix all causes of poor spectral quality.
The effective sampling time pattern of the instrument across targets or scan events. Duty cycle affects points per peak, sensitivity, and temporal resolution.
Profile data samples signal across the m/z peak shape; centroid data represents detected peaks by estimated positions and intensities. Processing choices affect which features remain and the values reported.
The expected distribution of isotope-related peaks for an ion. Useful for confirmation, charge assessment, and detecting overlaps or annotation errors.
The ability of a method to distinguish an analyte from interferences under stated conditions. Selectivity is built through sample prep, separation, acquisition, and review criteria together.
Checks that provide evidence the system meets specified criteria for the intended analysis. Passing does not guarantee every unknown sample is valid.
Software estimation of peak area or height, sometimes adjusted manually. Any manual reintegration should be rule-based, documented, and reviewable.
Gradual change in instrument behavior over time.
Detector regime where increased ion input no longer produces proportional response.
Techniques
Short technique orientations. These are not separate completed course tracks.
Volatile analytes, EI spectra, library matching, retention behavior, and contamination/carryover patterns that mimic real signal.
Liquid-phase separation, electrospray behavior, adduct formation, matrix effects, and batch-to-batch reproducibility.
Matrix selection, crystal quality, spot preparation, pulse timing, and spectrum variability caused by sample presentation.
Elemental analysis, isotopes, plasma tuning, interferences, internal standards, and washout-driven memory effects.
Start with the question and sample constraints: target type, volatility, polarity, expected concentration range, required selectivity, throughput, and what kind of evidence the final decision requires.
Users often overinterpret single peaks, ignore blanks, confuse tuning with calibration, and change multiple parameters at once. Record the evidence and isolate one hypothesis before changing a method.
Workflows
High-level decision checklists, not protocols. They highlight questions to ask; the instrument method and laboratory SOP supply the actual sequence and acceptance limits.
Typical path for targeted or semi-targeted analytes in complex matrices.
For volatile/semi-volatile analysis where separation quality and library interpretation both matter.
Focuses on calibration integrity, interference control, and long-batch stability.
Good workflows define stop points. If blanks fail, washout is poor, or QC trends drift, the right move is often to pause and diagnose, not to continue collecting invalid data.
Teach users to report what is measured, how it was confirmed, what thresholds were applied, and what limitations remain. This is where technical quality becomes operational trust.
Method Development
Reference notes for framing a method-development question. These do not replace an experimental plan, validated procedure or technique-specific guidance.
Start with what decision the data must support, not what the instrument can do.
Choose a small set of high-impact variables to screen first before fine tuning everything else.
A method is not complete when it works once. It is complete when it survives expected variability.
Not every lab needs formal DOE for every method, but every lab benefits from disciplined variable control. Interactions can make a structured design more informative than changing one factor at a time.
Pick objective metrics before optimization begins: signal stability, mass accuracy, points per peak, carryover threshold, QC recovery, precision, and false positive tolerance.
Data Analysis
Practical review prompts. Follow the linked worked primers for complete examples; other cards are brief reference notes.
Check acquisition metadata, instrument status flags, calibration state, chromatogram quality, and obvious contamination before trusting software output. This catches many problems earlier than peak table review.
Thresholds reduce noise but can also bias low abundance features. Centroiding, smoothing and deisotoping can each change the feature list; compare processing choices against suitable references and raw data.
Blank subtraction is useful, but it can conceal contamination trends or unstable systems if applied mechanically. Learn when to subtract, when to annotate, and when to stop the batch.
Use orthogonal evidence: exact mass (where relevant), isotope patterns, fragments, retention behavior, library score interpretation, and controls. No single metric should carry the full claim.
Evaluate calibration residuals, weighting assumptions, internal standard behavior, and reintegration decisions using documented criteria rather than visual preference alone.
Record processing version, peak integration edits, exclusions, and manual interventions. Auditable review protects both scientific quality and operational credibility.
QA / QC
Quality-control planning prompts. Frequencies, limits and required actions come from the applicable method and laboratory procedures.
Define a short, repeatable startup routine that answers whether the system is ready for real work.
Build QC placement and acceptance rules that detect drift without forcing constant reruns.
When methods or processing rules change, preserve comparability through explicit versioning and impact notes.
Trend a small set of meaningful metrics first: response, mass error, retention time, background, and blank carryover. The goal is early detection, not dashboard inflation.
Users often report precision from ideal runs only. State whether replicates vary by run, day, operator, instrument or laboratory; do not call all of these repeatability.